The Model is an Infusion Only syringe pump employing a microprocessor controlled
small step angle stepping motor driving a lead screw and half nut. A key pad is used
for data entry of flow rate and syringe diameter. The microprocessor calculates the
cross sectional area of the syringe to be used and calibrates the flow rate for that
syringe

Type Microprocessor multiple syringe, infusion or
infusion/withdrawal
Size (HxWxD) 140 x 220 x 220 cm (5.5 x 11 x 8.75 in)
Weight 10 lbs (4.5 kg)
Power 115/230 VAC, 50-60 Hz via selector switch, 45 W
Leakage to Ground Typically < 15 mA
Ground Resistance Typically < 0.2 ?
Voltage Operating Range 95 to 130 VAC, 50/60 Hz